The 4Pi-STED-microscope is the result of combining the two unrelated concepts of Stimulated emission depletion (STED) microscopy and 4Pi-microscopy. Here, a fluorescent sample is placed in the common focus of two opposing lenses, but excitation and detection are performed through a single lens (4Pi mode A). The green excitation pulse is immediately followed by a red STED-pulse, which enters the focal region through both lenses inducing stimulated emission of the excited fluorescent molecules to the ground state. To permit fluorescence emission from the center but suppress it from neighbouring regions it is useful to phase shift the STED beam to have a minimum at the center.

Property Value
dbo:abstract
  • The 4Pi-STED-microscope is the result of combining the two unrelated concepts of Stimulated emission depletion (STED) microscopy and 4Pi-microscopy. Here, a fluorescent sample is placed in the common focus of two opposing lenses, but excitation and detection are performed through a single lens (4Pi mode A). The green excitation pulse is immediately followed by a red STED-pulse, which enters the focal region through both lenses inducing stimulated emission of the excited fluorescent molecules to the ground state. To permit fluorescence emission from the center but suppress it from neighbouring regions it is useful to phase shift the STED beam to have a minimum at the center. (en)
dbo:wikiPageExternalLink
dbo:wikiPageID
  • 4739384 (xsd:integer)
dbo:wikiPageLength
  • 1014 (xsd:nonNegativeInteger)
dbo:wikiPageRevisionID
  • 1086623719 (xsd:integer)
dbo:wikiPageWikiLink
dbp:wikiPageUsesTemplate
dct:subject
gold:hypernym
rdf:type
rdfs:comment
  • The 4Pi-STED-microscope is the result of combining the two unrelated concepts of Stimulated emission depletion (STED) microscopy and 4Pi-microscopy. Here, a fluorescent sample is placed in the common focus of two opposing lenses, but excitation and detection are performed through a single lens (4Pi mode A). The green excitation pulse is immediately followed by a red STED-pulse, which enters the focal region through both lenses inducing stimulated emission of the excited fluorescent molecules to the ground state. To permit fluorescence emission from the center but suppress it from neighbouring regions it is useful to phase shift the STED beam to have a minimum at the center. (en)
rdfs:label
  • 4Pi STED microscopy (en)
owl:sameAs
prov:wasDerivedFrom
foaf:isPrimaryTopicOf
is dbo:wikiPageWikiLink of
is foaf:primaryTopic of
Powered by OpenLink Virtuoso    This material is Open Knowledge     W3C Semantic Web Technology     This material is Open Knowledge    Valid XHTML + RDFa
This content was extracted from Wikipedia and is licensed under the Creative Commons Attribution-ShareAlike 3.0 Unported License